Series S-3100
|
| |
The PDA spectrophotometers
with great value for money. |
|
-
|

|
|
|
(0.7 MB) |
The S-3100 Series
is the standard measurement instrument to satisfy such customer
requirements. By integrating hardware and sofrware from SCINCO,
a company with a proven history of innovation in PDA UV-Vis
instrumentation, the S-3100 Series makes the accuracy of a research-grade
spectrophotometer accessible to all laboratories. Offering a
standard interface configuration utilizing USB port to ensure
fast communication, the S-3100 Series facilitates your research.
In addition, the control of data by LabPro Plus Software enhances
the performance of SCINCO PDA UV-Vis Spectrophotometers. While
it is easy to learn and integrate with existing lab operations,
the S- 3100 Series also offers new-generation innovations and
ease-of-use features that improve the quality of results, while
boosting overall laboratory productivity.
- Dual
Lamp Source
Consists of an in-line Deuterium and tungsten dual lamp system
that maintains high-intensity light in the range of 190 to 1100
nm. It increases the energy intensity while utilizing a minimal
optical path. The "see-through" optical design gives increased
reliability and also provides a perfect, seamless transition
from UV to Visible wavelengths. Photometric accuracy is thus
assured, giving you maximum confidence in your results.
- Photodiode
Array (PDA) Detector The photodiode array(PDA) detector
provides an efficient analytical environment in which you can
obtain a reliable data through superior reproducibility, high
accuracy and an integrating function. It helps you to obtain
fast data acquisition and to be able to detect multi-wavelength
at the same time. It is also very useful for rapidly changing
samples. The 1024 channel PDA detector provides a spectral bandwidth
of 0.75nm (S-3130) or 0.95nm (S-3150).
- Flexibility
for Variable Inbterfaces
The S-3100 Series is designed to allow a higher stability through
RS-232C interface between spectrophotometer and the PC and the
rapid communications enable maximum data to be acquired. Thus,
the S-3100 Series can be effectively used in applications such
as Kinetics, enzyme activity, Protein research and DNA analysis,
etc.
- Modular Structure
The electronic and optical components of the S-3100 Series are
designed as a unique modular system. This modular design facilitates
easy maintenance and service.
- LabPro Plus Software
LabPro Plus Software is a powerful tool for collecting and processing
data in a wide variety of applications. LabPro Plus Software
allows you to easily acquire and analyze the highest quality
data from S-Series PDA UV-Vis Spectrophotometer. Instrument
validation is performed quickly and easily to quarantee best
results. LabPro Plus Software also provides diverse protein
methods and DNA Analysis in our Bio Package. Protein methods
include Bradford, BCA, Biuret, Lowry, TBA, and Direct UV. LabPro
Plus Software is compatible with windows 98/ME/2000 or XP. Built-in
data conversions and formats provided efficiently support Lab
Pro and LabPro Easy files, CSV, XLS, BMP and WMF. LabPro Plus
data is easily exported into Excel and other Windows programs.
- Specification
- Model : S-3130
(S-3150)
- Sources :
Pre-aligned Tungsten & Deuterium Lamp
- Lamp Detector
: 1024 channel PDA
- Wavelength
Range : 190 ~ 850 nm (190 ~ 1100 nm)
- Wavelength
Reproducibility : ±0.02 nm
- Wavelength
Resolution : 0.75 nm (0.95 nm)
- Data Acquisition
Mode : AU, %T, %R, Intensity, Concentration
- Photometric
Range : -3.0 ~ 3.0 AU
- Scan Speed
: 0.02 sec for full wavelength range
- Integration
Time : 0.02 ~ 6 sec
- Dimension
: 52.1 (W) ×23.1 (H) ×45.0 (D) cm, 14 kg
- Power Voltage
ranges : 110 or 220 V ±10 %, free of spikes Current
- Environmental
Operating temperature : 5°C~40°C
- Humidity :
less than 95 % relative, non-condensing.
-
Use
Identification, Multi-Component Analysis, Quantitative Analysis,
Kinetic Experiment, Protein Quantification, DNA Analysis.
|
|
| |
|
|
SO-S3130
|
SO-S3150
|
|
reference
|
| |
|
|
S-3130
|
S-3150
|
|
name |
| |
|
|
190 - 850
|
190 - 1100
|
(nm)
|
wavelength
|
| |
|
|
0.01
|
0.01
|
(nm)
|
resolution
|
| |
|
|
0.75
|
0.95
|
(nm)
|
bandwith
|
| |
|
|
± 0.02
|
± 0.02
|
(nm)
|
reproducibility
|
| |
|
|
-3.0 - +3.0
|
-3.0 - +3.0
|
(AU)
|
photometic range
|
| |
|
|
AU,
%T, %R, Intensity, Concentration
|
AU,
%T, %R, Intensity, Concentration
|
( - )
|
units |
| |
|
|
< 0.05*
|
< 0.05*
|
(%T)
|
stray light *@340 & 220nm
|
| |
|
|
0.02**
|
0.02**
|
(s)
|
scan speed *for
full wavelength range |
| |
|
|
yes
|
yes
|
( - )
|
RS-232 |
| |
|
|
|
|
|
|
|
|
|
|
|
|
(CHF)
|
price
|
| |
|
|
|
|
|
only!
|
| |
Communication
cable (RS-232/USB), mains lead, cell holder and instruction manual. |
|
included |
| |
Option
accessories: Peltier temperature
controller, cells & cell holders, auto sampler and sipper system,
refrigerated circulator, LabPro Plus software, transmission dip
probe and many others. |
|
|
|